Why 98% and Not 99% or 95%?
The 98% threshold is a convention rather than a law of chemistry: researchers, contract laboratories and suppliers tend to treat it as the lowest purity at which a peptide is fit for most research work. Below it, the impurities begin to distort experimental results and biological activity measurements, and for any compound that ends up in a human body they undermine the safety and predictability of what is taken.
Some applications require higher purity: peptides intended for in vivo work or clinical trial supply may require 99%+ purity with additional identity confirmation. Some applications may tolerate lower purity in specific analytical contexts. The 98% threshold represents the accepted floor for what the research community considers a credible general-purpose standard.
The figure only means something alongside the method that produced it. Different analytical methods can give different purity values for the same sample, so a purity stated without a methodology, or measured by something other than HPLC, cannot be compared directly with the 98% benchmark, which comes from RP-HPLC.
HPLC — High-Performance Liquid Chromatography
HPLC: what it measures
Purpose: Measures the relative proportions of compounds in a sample and reports purity as a percentage.
What it does not confirm: HPLC does not confirm molecular identity. A highly pure sample of the wrong compound will produce a high purity result by HPLC.
Standard variant for peptides: RP-HPLC (Reversed-Phase HPLC) using a C18 column, UV detection at 214–220 nm.
High-Performance Liquid Chromatography works by dissolving the sample in a liquid solvent (the mobile phase) and pumping it under high pressure through a column packed with a solid material (the stationary phase). Different components of the sample interact differently with the stationary phase: some move through the column quickly, others are retained longer. This is chromatographic separation, and it produces a series of peaks as components leave the column at different times.
A UV detector at the column exit measures the absorbance of each peak. For peptides, the standard detection wavelength is 214 nm (where the peptide bond absorbs light) or 220 nm. The relative area of each peak, expressed as a proportion of total peak area, gives the purity percentage. A sample with a single large peak at 98.5% of total peak area, with the remaining 1.5% distributed across minor impurity peaks, would report as 98.5% pure by HPLC.
What HPLC Cannot Tell You
HPLC cannot confirm the identity of the major peak. If the largest peak in a chromatogram belongs to a different compound from the one on the label, HPLC still reports it as highly pure, because it only compares peak areas.
That gap is what LC-MS identity confirmation closes.
LC-MS — Liquid Chromatography–Mass Spectrometry
LC-MS: what it measures
Purpose: Confirms molecular identity by measuring mass-to-charge ratio (m/z). Determines whether the compound is what the label claims.
What it does not replace: LC-MS confirms identity but does not replace HPLC for purity determination. It detects the major compound but is less suited to precise quantification of minor impurities.
Key output: Observed molecular weight compared against theoretical molecular weight for the stated CAS-registered compound.
Liquid Chromatography–Mass Spectrometry combines the separation capability of HPLC (the LC component) with the mass measurement capability of a mass spectrometer (the MS component). After chromatographic separation, the eluent from the column passes into an ionisation source, which converts the compound molecules into gas-phase ions. These ions are then separated in the mass analyser according to their mass-to-charge ratio (m/z) and detected.
The result is a mass spectrum, a plot of ion abundance against m/z, which gives the molecular weight of the compound. For a synthetic peptide, the observed molecular weight should match the theoretical weight calculated from the amino acid sequence, within the tolerance the laboratory states for its instrument; high-resolution instruments allow a much tighter match than a basic single-quadrupole one.
Interpreting the LC-MS Result
When reading an LC-MS result on a peptide COA, look for the following:
- Theoretical molecular weight: The calculated mass for the stated compound, derived from its amino acid sequence and the molecular formula. This should be stated on the COA alongside the CAS number.
- Observed (found) molecular weight: The mass measured by the instrument for the main peak in the LC-MS analysis.
- Delta or difference: Some COAs report the difference between theoretical and observed mass. A small delta, within instrument tolerance, supports the stated identity; a large one points to a different or modified compound.
- Multiple charge states: Larger peptides are often detected at multiple charge states (e.g., [M+2H]²⁺, [M+3H]³⁺). A credible COA from a well-equipped laboratory will show the relevant charge states and the calculated monoisotopic or average mass deconvoluted from these peaks.
Why Both Methods Are Required Together
Asking for both HPLC and LC-MS is not duplication, because each answers a different question. HPLC quantifies the main component against its impurities. LC-MS checks that the main component has the molecular mass of the claimed compound.
A COA showing only HPLC purity of 99.1% provides strong evidence of a highly pure sample, but no confirmation of what that sample is. A COA showing only LC-MS identity confirmation provides evidence of the correct molecular weight, but no quantification of how much of the sample that compound represents versus impurities.
Together they show that the sample is the stated compound (LC-MS) and that it is at least 98% pure (HPLC). A COA with only one of the two is incomplete.
Common Analytical Deficiencies in Grey-Market COAs
Knowing the methods makes it easier to spot weak paperwork. Common deficiencies:
- Purity reported by UV absorbance at a single point, without column chromatography, which is not HPLC
- HPLC reported without stating the column type, mobile phase, or detection wavelength, so the method cannot be assessed
- No batch number, or one that does not match the vial label, so the result cannot be tied to the product in hand
- No analysis date, or a date long before the sale; peptides can degrade in storage, so an old result may not describe the vial you receive
- LC-MS data absent, or present but showing only a single mass peak with no charge-state analysis for larger peptides
- Mass reported only as nominal molecular weight without comparison to theoretical value
- COA produced using instrument software that is identifiable as consumer-grade or non-laboratory equipment
None of these proves the product is poor, but each means the documentation falls short of the evidence a 98% HPLC / LC-MS benchmark implies.
The Broader Verification Picture
HPLC purity and LC-MS identity are essential analytical checks, but they are not the complete picture. They do not address bacterial endotoxin content, residual solvents, sterility (for injectable compounds), or peptide folding and secondary structure in more complex cases. For research purposes, these additional tests may or may not be relevant depending on the application.
For anyone researching peptides in the UK, HPLC plus LC-MS is the practical minimum documentary evidence of quality. Laboratory paperwork says nothing about legal status. Research peptides without a UK marketing authorisation are unlicensed medicines when intended for human use; the MHRA position on the compound and a conversation with a licensed healthcare professional come separately, because these compounds sit outside the UK licensed medicines framework. See our summary of what a London buyer should check before purchasing for where these methods fit in.